Lesson 19: Medical Physics (Option)
🎧 Audio narration for this lesson is coming soon.
Medical physics applies waves, fields, and quantum physics to the human body — imaging the invisible and treating the untreatable. This A-level option covers X-ray imaging (attenuation and contrast), CT scanning (reconstruction from projections), ultrasound (reflection and Doppler blood-flow measurement), MRI (nuclear magnetic resonance), and the physics of vision and hearing. Every technique is core physics pointed at the body: exponential attenuation, wave reflection, resonance, and the Doppler effect saving lives daily.
X-rays — Seeing Through the Body
X-rays (0.01–10 nm, keV–MeV photons) attenuate exponentially: I = I₀e^{−μx} — μ the attenuation coefficient (depends strongly on Z — photoelectric absorption ∝ Z³ roughly — bone (Ca, Z=20) absorbs far more than soft tissue). Half-thickness x½ = ln2/μ. Contrast: bone vs tissue differ in μ — the shadow image. Worked: 50 keV X-rays, μ_bone = 0.8 cm⁻¹, μ_tissue = 0.2 cm⁻¹. Through 10 cm tissue + 2 cm bone: I/I₀ = e^{−(0.2×10 + 0.8×2)} = e^{−3.6} ≈ 0.027; through 12 cm tissue: e^{−2.4} ≈ 0.091 — bone shadows are 3× darker. Contrast agents: barium (Z=56) meals coat the gut; iodine (Z=53) injected for blood vessels — high-Z, high-μ, temporary.
X-ray production: bremsstrahlung (electrons decelerated in tungsten target — continuous spectrum) + characteristic lines (K-shell knocked out, L→K transition emits Kα — tungsten 59 keV). Tube voltage sets max photon energy (E_max = eV — 100 kV → 100 keV photons); current sets intensity (more electrons). Filters (aluminium) remove soft X-rays (absorbed by skin, contributing dose without image) — beam hardening as a safety feature. Rotating anodes spread the heat (kW of electron beam on mm²).
Dose and risk: absorbed dose (gray = J/kg), equivalent dose (sievert = gray × radiation weighting factor — alpha 20, gamma 1), effective dose (tissue weighting). Chest X-ray ~0.02 mSv; CT abdomen ~8 mSv; annual background ~2.7 mSv (UK — radon-heavy Cornwall higher). ALARP governs medical exposures: benefit must outweigh risk (a CT finding a tumour outweighs 8 mSv). Lead aprons, collimation (narrow beams), and digital detectors (less dose than film) minimise exposure.
Worked — attenuation thickness: a beam must be reduced to 10% of its incident intensity passing through tissue with μ = 0.35 cm⁻¹. From I/I₀ = e^{−μx} = 0.10: x = ln(10)/0.35 = 2.303/0.35 ≈ 6.6 cm. In half-thicknesses: x½ = ln2/0.35 ≈ 1.98 cm, and 10% ≈ (½)^3.32, so 3.32 × 1.98 ≈ 6.6 cm ✓. Fluency both ways (exponent and half-thickness counting) is what exam speed looks like.
CT Scanning — Reconstruction
CT: an X-ray tube rotates around the patient, detectors measuring attenuation along hundreds of angles; computation reconstructs the 2D slice (Radon transform — the mathematics of projections). Each voxel's μ mapped to Hounsfield units: HU = 1000(μ − μ_water)/μ_water — water 0, air −1000, bone +1000+, blood ~+40. Windowing displays tissues: brain windows, bone windows — the radiologist's contrast controls. A chest CT is ~500 slices — millimetre anatomy in 3D.
Worked — Hounsfield units: a tissue has μ = 0.21 cm⁻¹ where water has μ_water = 0.20 cm⁻¹. HU = 1000(0.21 − 0.20)/0.20 = +50 — soft-tissue range (muscle and many tumours live at +20 to +80; fat is negative, around −50 to −100). The HU scale converts a raw attenuation measurement into a diagnostic number with water as the universal reference — calibrate once, compare everywhere.
Worked concept: two projections of a 2×2 grid suffice to illustrate — horizontal beams give row sums, vertical give column sums; solving the equations yields the four pixels (with ambiguities resolved by more angles). Real CT uses filtered back-projection over 1000+ angles — the algorithm is applied mathematics (Fourier methods). Spiral CT: continuous rotation + table movement — whole-body scans in seconds; cardiac CT gates to the heartbeat (ECG-triggered acquisition — freezing motion). Contrast with plain X-ray: CT separates overlapping structures (slices vs shadows) and quantifies μ (Hounsfield numbers diagnose: fresh blood vs old, tumour vs cyst). Cost: higher dose (more projections) and computation. PET-CT fuses metabolic (Lesson 16's positrons) with anatomical imaging — cancer staging's gold standard. The option's techniques combine: physics is modular, medicine uses the modules together.
Ultrasound — Sound as a Probe
Ultrasound (> 20 kHz; medical 1–15 MHz): generated by piezoelectric crystals (AC voltage → mechanical vibration — the reverse of the microphone effect). Wavelength in tissue (v ≈ 1540 m/s): at 5 MHz, λ = 0.31 mm — millimetre resolution (resolution ~λ). Reflection at boundaries: fraction reflected depends on acoustic-impedance mismatch Z = ρv — soft-tissue/bone reflects strongly (bright), tissue/tissue weakly. Time of flight gives depth: d = vt/2 (there and back) — the A-scan principle; B-scan sweeps to build 2D images.
Worked: echo returns 0.13 ms after the pulse: d = 1540(0.13×10⁻³)/2 ≈ 0.10 m = 10 cm deep. Gel coupling: air's Z mismatches skin hugely (99.9% reflected) — gel matches impedances, letting sound in. Frequency trade-off: higher f → better resolution but more attenuation (α ∝ f — high frequencies absorbed) — obstetrics uses 3–5 MHz (deep), eyes/thyroid 10–15 MHz (shallow detail). No ionising radiation — safe in pregnancy (the decisive advantage over X-ray/CT).
Doppler ultrasound: moving blood shifts the reflected frequency — Δf = 2f₀v cosθ/c (factor 2: the blood acts as moving observer AND moving source). Worked: 5 MHz probe, blood at 0.5 m/s toward probe (θ = 0): Δf = 2(5×10⁶)(0.5)/1540 ≈ 3.25 kHz — audible! (Doppler fetal monitors make it audible deliberately.) Worked — angled probe: a 7.5 MHz probe views a vessel at θ = 60° with blood speed 0.35 m/s: Δf = 2(7.5×10⁶)(0.35)(cos60°)/1540 = 2.63×10⁶/1540 ≈ 1,705 Hz ≈ 1.7 kHz. The cosθ factor is the clinical trap: near θ = 90° the shift vanishes — sonographers keep θ ≤ 60° so the correction stays reliable. Colour Doppler maps velocity (red toward, blue away — conventions) — cardiologists watch valves leak in real time. The Doppler effect (Lesson 6) as a blood-flow meter.
MRI — Resonance Imaging
Protons (hydrogen nuclei) are magnetic dipoles — in a strong B₀ field (1.5–3 T, superconducting magnets — Lesson 11) they align parallel/antiparallel (tiny excess parallel — Boltzmann). Larmor precession: spins precess at f = γB₀/2π (γ/2π = 42.6 MHz/T for protons) — at 1.5 T: 63.9 MHz (radiofrequency). Worked: at 3.0 T, f = 42.6 × 3.0 ≈ 128 MHz — FM-radio frequencies, which is why MRI suites are Faraday-caged against radio interference. RF pulse at the Larmor frequency tips spins (resonance — like pushing a swing); they re-emit RF as they relax — the signal. Different tissues relax at different rates (T1: realignment; T2: dephasing) — the contrast mechanism (fat vs water vs tumour distinct).
Spatial encoding: gradient coils make B vary with position — Larmor frequency becomes position-dependent (frequency encodes x, phase encodes y) — Fourier transform reconstructs the image (mathematics again!). No ionising radiation (RF is non-ionising) — safe for repeated scans; but: projectile risk (ferromagnetic objects fly into the bore — screening essential), claustrophobia, loud gradients, and cost (£1M+ scanners). Functional MRI: blood-oxygen changes alter T2* (BOLD signal) — mapping brain activity (neuroscience's workhorse).
Comparing modalities: X-ray (cheap, fast, ionising, 2D shadows); CT (3D, quantitative, higher dose); ultrasound (safe, real-time, operator-dependent, poor through bone/air); MRI (superb soft-tissue contrast, safe, slow, expensive, metal restrictions); PET (metabolic, radioactive tracer, fused with CT). 'Choose the modality' exam questions: match the physics to the clinical need (pregnancy → ultrasound; brain tumour detail → MRI; trauma bleed → CT speed; fracture → X-ray).
Vision, Hearing, and Exam Technique
The eye: cornea + lens focus (converging system, ~60 dioptres total) onto the retina; accommodation (ciliary muscles change lens shape); defects — myopia (eyeball too long, corrected with diverging lenses), hyperopia (too short, converging), astigmatism (cylindrical). The ear: pinna → canal (resonance ~3 kHz) → eardrum → ossicles (lever + area ratio amplifying pressure ~20×) → cochlea (frequency-separated hair cells — place theory). Hearing loss: conductive (mechanical) vs sensorineural (nerve) — diagnosed by comparing air/bone conduction (Rinne/Weber tests — physics in the clinic).
Exam technique: (1) attenuation questions — write I = I₀e^{−μx}, identify μ for each material, sum exponents for layers; (2) ultrasound — d = vt/2 (the half is the mark), Doppler with the factor 2 derived; (3) MRI — Larmor frequency from f = (γ/2π)B₀ = 42.6 MHz/T × B₀, resonance condition stated, T1/T2 as the contrast source; (4) 'compare techniques' — table format: physical principle, resolution, safety, cost, best use — examiners reward structured comparison over prose; (5) dose questions — gray vs sievert distinguished, ALARP invoked, background compared (mSv scale familiarity: background 2.7, chest X-ray 0.02, CT 8).
Medical physics is applied core physics — attenuation (Lesson 15's gamma), waves (Lessons 6–7), resonance (Lesson 6), magnetism (Lesson 11), quantum (Lesson 8's photons). The option tests whether you can recognise old physics in new clothes — the ultimate synoptic skill.
Common mistakes
- Forgetting the factor 2 in ultrasound ranging (pulse goes there AND back) and in Doppler (double shift). Both factors are the commonest lost marks in this option.
- Gray vs sievert: gray is absorbed energy per kg; sievert folds in radiation type and tissue sensitivity. A dose in 'gray' cannot be compared with background (quoted in mSv) until weighted.
- μ units and layer sums: exponents add layer by layer with each material's own μ and thickness — a single average μ for mixed tissue/bone is wrong.
- MRI 'radiation' confusion: RF photons are non-ionising (MHz, ~10⁻⁷ eV each); MRI's hazards are projectiles, implants and heating — not dose. Stating MRI is 'radiation-free' without qualification also loses the nuance.
- Larmor slip: f = (γ/2π)B₀ with γ/2π = 42.6 MHz/T — don't divide by 2π twice.
Exam Experience
(a) A beam passes through tissue with μ = 0.35 cm⁻¹. Calculate the thickness that reduces the intensity to 10% of its incident value. [3] (b) A tissue has μ = 0.21 cm⁻¹; for water μ = 0.20 cm⁻¹. Calculate its Hounsfield unit value and comment on the tissue type. [3] (c) A 7.5 MHz ultrasound probe views blood flowing at 0.35 m/s at 60° to the beam (speed of sound in tissue 1540 m/s). Calculate the Doppler shift. [3] (d) Calculate the proton Larmor frequency in a 3.0 T MRI scanner (γ/2π = 42.6 MHz/T). [2] (e) State, with a physics justification, the most suitable imaging method for routine imaging of a fetus. [2]
Mark scheme (a) 0.10 = e^{−μx} → x = ln10/0.35 = 2.303/0.35 ≈ 6.6 cm. [3] (b) HU = 1000(0.21 − 0.20)/0.20 = +50 [2]; soft-tissue range (e.g. muscle) — far below bone (+1000), above fat (negative). [1] (c) Δf = 2f₀v cosθ/c = 2 × 7.5×10⁶ × 0.35 × cos60°/1540 ≈ 1.7×10³ Hz. [3 — factor 2 AND cosθ both required] (d) f = 42.6 × 3.0 ≈ 128 MHz. [2] (e) Ultrasound [1]; it uses non-ionising sound (no dose to fetus/mother), with real-time imaging — unlike X-ray/CT (ionising) it is safe for repeated pregnancy scans. [1]
🔬 Interactive Diagram: "Medical Physics (Option)"
The key components of this lesson, mapped around its central idea.
🗺️ Interactive: Key Concepts Map
Interactive — hover or tap a concept to see its definition.
Key Terms
- Attenuation
- I = I₀e^{-μx} — exponential absorption; μ ∝ Z³ (photoelectric).
- Hounsfield unit
- HU = 1000(μ-μ_water)/μ_water — CT's quantitative tissue scale.
- CT reconstruction
- Multiple projections → computed slices (Radon transform, back-projection).
- Piezoelectric effect
- Crystal converting voltage ↔ vibration — ultrasound generation/detection.
- Acoustic impedance
- Z = ρv — mismatch causes reflection; gel matches skin.
- Doppler ultrasound
- Δf = 2f₀v cosθ/c — blood velocity from frequency shift.
- Larmor frequency
- f = γB₀/2π (42.6 MHz/T for protons) — MRI's resonance condition.
- T1/T2 relaxation
- Spin realignment/dephasing rates — MRI's tissue-contrast mechanism.
- Absorbed/equivalent dose
- Gray (J/kg) vs sievert (× radiation weighting) — dose vs biological harm.
- ALARP
- Benefit must outweigh dose — medical exposure justification.
Discussion Questions
- 50 keV X-rays (μ_tissue = 0.2 cm⁻¹, μ_bone = 0.9 cm⁻¹) image a wrist: 8 cm tissue alone vs 8 cm tissue + 1.5 cm bone. Compute transmitted intensities and the contrast ratio. Explain why iodine (Z=53) is injected for angiography, with reference to μ's Z-dependence.
- Derive d = vt/2 for ultrasound ranging. A 7.5 MHz probe images a structure with echo delay 78 μs (v = 1540 m/s). Find its depth and the wavelength-limited resolution. Explain the frequency trade-off for imaging a deep abdominal organ vs the thyroid.
- Derive the Doppler-shift formula for ultrasound (factor 2 explained). Blood flows at 0.8 m/s at 60° to a 4 MHz beam. Compute Δf and state whether it is a rise or fall for blood approaching the probe. How does colour Doppler use this?
- A 1.5 T MRI scanner: compute the proton Larmor frequency. Explain (a) why the main field must be so strong, (b) how gradients encode position, (c) what T1 and T2 contrast reveal, and (d) two safety hazards with their physical origins.
- Compare X-ray, CT, ultrasound, and MRI for: (i) suspected skull fracture, (ii) 20-week pregnancy check, (iii) brain tumour characterisation, (iv) emergency abdominal bleeding. Justify each choice with physics (not just 'it is used for this').